Measurement Specialties Inc. Model 9116 User’s Manual
Page 59 www.MEAS-SPEC.com
data selected for inclusion in each stream packet
Enable Valve Position Status (reserved for future use)
Enable DH Temperature Status (see bit map below)
Enable Pressure EU Data (default if “05” never executed after “00”)
Enable Pressure A/D Counts
Enable DH Temperature EU Data (degrees C)
Enable DH Temperature A/D Counts
Enable DH Temperature Voltages
Any DH Temperature Status datum is delivered as a two-byte binary bit map (16-bit, big
endian) with each bit representing the status of DH #16 through DH #1 respectively. A bit value
of 0 (zero) indicates the DH is operating within its specified limits. A value of 1 (one) indicates
the DH is outside its specified limits.
The above example indicates that Channels 1 and 16 are operating outside the specified
temperature limits.
** This status field (0001) cannot be specified for Model 9116.
However it is shown should the capability be added to future
firmware versions. Currently, only Models 9816 and 903x can return
Valve Position status in their streams.